Clock Recovery and Rehab
by Tom Lecklider, Senior Technical Editor
The high-speed serial interface signals supporting today’s computer and
communications systems are too fast for most general-purpose test equipment. For
example, PCI Express operates at a 2.5-GHz bit rate, equivalent to a 1.25-GHz
fundamental frequency. For accurate measurements, oscilloscope manufacturers
recommend that a scope’s bandwidth should be at least as high as a digital
signal’s fifth harmonic or 6.25 GHz in this case.
Although the average scope bandwidth has been increasing over the years, today
it’s far closer to 1 GHz than 6.25. And, PCI Express already has been surpassed
by PCI Express 2.0 at 5 Gb/s, fully buffered DIMM running at 4.8 Gb/s, and fibre
channel at 4.25 Gb/s. In all three of these examples, bandwidth greater than 10
GHz is required to include the fifth harmonic of the fundamental frequency.
Recently, major scope manufacturers have introduced instruments with bandwidths
from 15 GHz to 20 GHz. Nevertheless, it’s unlikely that a huge used-equipment
market suddenly will develop based on thousands of abandoned lower bandwidth
scopes.
A more probable outcome is that engineers will use high-bandwidth sampling
scopes. The new DSOs only are necessary if single-shot fast transients must be
investigated. Otherwise, most serial interface characterization can be
accomplished with repetitive sampling scopes.
Sampling Scopes
Traditional sampling scopes require a trigger that is synchronous with a
repetitive signal. At each successive trigger, the scope samples the signal at a
slightly later time. For example, if you wanted to examine a 1-ns long portion
of a signal and your scope could display 1,000 points horizontally, each sample
would be delayed an additional 1 ps from the trigger as shown in Figure 1. In
this way, a very detailed view of the signal is built up, equivalent to an
extremely high sampling rate, much higher than even that of the newest 50-GS/s
DSO.
|

Figure 1. Sampling Process for a Pattern Waveform
Courtesy of
Agilent Technologies |
Sampling scopes boast bandwidths up to 80 GHz or more and 14-b vertical
resolution. The trade-off made to achieve these impressive specs is a sampling
rate typically less than 1 MHz. It’s important to understand that the built-up
picture is not an expanded version of a single-shot DSO acquisition: Rather, it
contains elements of thousands of signal repetitions. This method of operation
results in many subtle implications. Nevertheless, if you can work with a
repetitive signal, you may have access to much greater fidelity in time and
amplitude than achievable by other means.
Triggering
There is a finite time delay between the instant a trigger occurs at the scope’s
front-panel connector and when the corresponding sample is taken. For example,
in the Agilent Technologies Model 86100C Infiniium Digital Communications
Analyzer (DCA), this delay is approximately 24 ns. Here the word approximately
means that the minimum trigger-to-sample delay is not exactly 24.000 ns although
its actual value has less than 100-fs uncertainty. LeCroy lists a 12-ns delay
for its WaveExpert® Series of sampling scopes.
Variation in the trigger timing relative to the input signal translates directly
into jitter. The 86100C has a fastest time-base setting of 2 ps/div or 500
fs/div with the 86107A precision time-base reference module. At these speeds, it
can be very difficult to maintain low jitter using a separate trigger signal. To
put things in perspective, an electrical signal in a typical coaxial cable
travels about 0.025 mm in 100 fs. This distance is equivalent to the change in
length of a 1-m cable caused by a 2°C temperature change.
Obviously, sampling scopes are successfully used with separate trigger sources.
The point is simply that accurately measuring subpicosecond effects involves
many types of errors that usually are ignored for nanosecond-scale signals.
Unless you can account for these errors or in some way minimize them, few digits
of the high-precision measurements your scope displays will have significance.
Some sampling scopes can develop a pattern-based trigger internally. For
example, the 86100C enhanced trigger option provides the PatternLock facility.
PatternLock automatically detects the pattern length, data rate, and clock rate
and produces a stable trigger at a user-selectable position within the pattern.
This feature has required changes to the traditional sampling-scope architecture
that provide additional related benefits. For example, by taking samples
sequentially through a data pattern, the time relationship among a group of
samples is maintained. This means that should an anomaly be found in an eye
diagram, the effects of specific bit transitions can be investigated.
The Tektronix Digital Serial Analyzer Sampling Oscilloscope Model DSA8200, when
fitted with the Model 80A06 PatternSync Module, also can develop a stable
trigger from a repeating data pattern. The 80A06 is a programmable counter that
accepts clock rates from 150 MHz to 12.5 GHz. It divides the clock rate by a
number from two to 223 to produce a stable trigger synchronized to the data
pattern.
By using the 80A06 together with the FrameScan™ acquisition mode, you can
repetitively acquire selected bits within the data pattern. FrameScan
automatically advances the trigger location to the next bit position after the
acquisition process has been completed. Alternatively, you can elect to capture
data relating to only one bit position. FrameScan data can be averaged.
One complete data pattern is termed a frame and a bit within the pattern a
subframe. To acquire successive subframes, the scope’s time base must be set to
units of bits rather than time. In addition, you must enter the data rate,
pattern length, and data-to-clock ratio or select the corresponding auto-sync
options so that the module can automatically detect these parameter values.
Because there is no unique beginning to a repeating pattern, you must enter a
relative trigger bit value to advance or retard the effective trigger point
within the pattern.
LeCroy’s WaveExpert Model NRO 9000 offers conventional sequential sampling as
well as coherent interleaved sampling (CIS). Because CIS is a new acquisition
mode that operates differently than both a traditional sampling scope and a DSO,
LeCroy coined the term near real-time oscilloscope (NRO). Sampling scopes
generally have very short acquisition memories when compared to the megabytes of
DSOs. With CIS, sampling rates are from one to two orders of magnitude faster,
and longer memories are more practical.
Traditional coherent sampling establishes a rate relative to the repetition rate
of the data pattern (Figure 2a). For example, if a data pattern repeated every
eight bits or unit intervals (UIs), the display required 1,000 points, and all
eight UIs were to be shown, 125 samples per UI would be needed. The incremental
delay between sampling points would be adjusted so the successive positions of
the 1,000 samples progressed evenly through the eight UI data pattern.
|

Figure 2a. Coherent Sampling Data Acquired by Locking to the Data
Pattern Rate |
CIS works in the same way as coherent sampling but subdivides the data pattern
to achieve a higher sampling rate as shown in Figure 2b. Because of this
subdivision, successive samples no longer are in the correct time order and must
be re-ordered post-acquisition.
|

Figure 2b. CIS Acquired Samples at a Fixed Rate by Dividing the Data
Pattern Into Segments and Reordering Samples With Respect to Their
Position in the Segment
Courtesy of
LeCroy |
CIS and coherent sampling use a PLL to synchronize the sampling clock rate to
the data rate. This is a different mode of operation than FrameScan or
PatternLock that reference the positions of successive samples to a pattern
trigger. CIS is controlled by a samples/UI parameter while FrameScan and
PatternLock acquisitions run at fixed, selectable time-base rates.
A general consideration that applies to all sampling scopes concerns using
subrate or divided clocks to generate eye diagrams. It is common for clocks to
run at some integer division of the data rate. For example, a 40-Gb/s data
stream may be developed by multiplexing four 10-Gb/s streams so the available
clock rate would be 10 GS/s.
If the clock is used as the trigger, an eye diagram will be formed that shows
all possible combinations of data transitions. At least, that’s the objective.
However, if the length of a repetitive data pattern is directly related to the
subrate clock, only a fraction of the pattern may be sampled. In the 40-Gb/s
data and 10-GS/s clock example, if the pattern is 20 b long, the same 5 b will
be sampled by the 10-GS/s clock. Only 25% of the data transitions will make up
the eye diagram. To avoid this problem, the pattern length must not be evenly
divisible by the clock division factor, four in this case.
Clock Recovery
One way that clock-to-data skew and uncertainty have been controlled is to embed
the clock within the data stream. There is only one signal so the clock-to-data
skew is, by definition, zero. On the other hand, the clock must be recovered
from the data stream, usually by using a PLL (Figure 3).
|

Figure 3. Clock Recovery Module Block Diagram
Courtesy of
Agilent Technologies |
In theory, it makes sense to recover the clock using a PLL with characteristics
similar to those of the actual clock/data receiver (CDR) that will be used in
the system. By matching the PLL to the CDR loop bandwidth and related parameters
such as rate of pass-band roll off, the recovered clock will have the same
amount of jitter as the clock that will be used in the real system.
Nevertheless, there are reasons to independently control the instrumentation PLL
bandwidth.
For example, if the loop bandwidth is very high, the loop will follow even
high-frequency jitter so the recovered clock will have most of the same jitter
that is affecting the data. If you trigger a sampling scope with this type of
recovered clock, because jitter is causing the trigger and data to move by the
same amount and in the same direction, the acquired signal will be very clean.
This mode of clock recovery would be useful for examining rise/fall times and
pulse widths without the added uncertainty caused by jitter.
On the other hand, a very narrow loop bandwidth causes the PLL to ignore most of
the data jitter, producing a very stable clock. In this case, almost all the
data jitter will be apparent in the acquired signal because there is little
clock jitter. When viewed as part of the overall signal-acquisition process,
clock recovery PLL loop bandwidth has a converse effect: A wide loop bandwidth
acts like a low-pass narrowband acquisition filter while a narrow loop bandwidth
acts like a wideband acquisition filter.
An example of this type of PLL effect is found in LeCroy’s description of the
CDR-E135 Electrical Clock Recovery Module. With the extended frequency range
option, data rates from 622 Mb/s to 13.5 Gb/s are accepted. The unit is
specified with a 6-MHz PLL loop bandwidth and recovers a clock with typically no
more than 500-fs rms jitter.
The important point is a data sheet comment about the action of the CIS PLL when
used with this module: “The recovered clock is used by the [WaveExpert] time
base as a reference, and the action of the PLL in the CIS time base removes any
residual jitter in the recovered clock, allowing the time-base jitter to be as
low as 250 fs rms when the high-stability CIS time-base option is present in the
oscilloscope (HCIS).”
The data pattern transition density also affects PLL gain and loop bandwidth.
Transition density is defined as the ratio of the number of transitions that
actually take place to the number that could. A 010101 pattern is changing on
every bit position so it has a 100% transition density. A 001100 pattern has
only a 50% transition density. This effect can be important, and some standards
specify the pattern used for testing so the test is repeatable.
Many serial data specifications call for the use of a golden PLL, one in which
the loop bandwidth is matched to that of the standard. For example, a multirate
clock recovery option is available for Wavecrest SIA instruments. For a wide
range of serial communications standards with bit rates from 1.0625 Gb/s to 4.25
Gb/s, the bandwidth of this option is the bit rate divided by 1,667. It also is
common for some standards to specify bit rate/2,500. In these cases, the PLL
loop bandwidth is set by the data rate so you are limited to acquiring data
within the parameters of particular standards.
Agilent’s Model 83496A/B Series of clock recovery modules features a continuous
range of input data rates from 50 Mb/s to 13.5 Gb/s as well as continuously
adjustable loop bandwidth from 15 kHz to 10 MHz. The Tektronix Model 80A07
Electrical Clock Recovery Module provides a large number of clock division
ratios between 1 to 648, continuously variable data rates from 100 Mb/s to 12.5
Gb/s, and continuously tunable loop bandwidth from 100 kHz to 12 MHz. Either
module can function as a golden PLL.
Enhanced Accuracy Clocking
Time-base jitter less than 1 ps is impressive, but can you achieve better? Yes
you can, and reference phase measurement is the key. This technique recently was
recommended by the ITU in document O.172 as an accurate way to measure jitter.
Although not new, the approach has benefited from improved components and
implementation.
A Tektronix application note describes the operation of the Model 82A04
Phase-Reference Module.1 At the heart of the module are two samplers that are
driven simultaneously. One samples the input data and the other a user-supplied
synchronous clock. Typically, to achieve the lowest possible uncertainty, this
clock would be generated by a highly stable, low-noise reference oscillator such
as an RF synthesizer.
Before the module can be used to acquire signals, the reference clock must be
characterized. A total of 250,000 clock cycles are sampled by a free-running
asynchronous strobe and the information used to build a voltage-to-phase lookup
table. Using this information, the module can convert a sample of the reference
clock to an equivalent clock phase. Ideally, the clock should be a perfect
sinusoid, but the module is designed to cope with distorted clocks more often
seen in practice. Given a user-supplied reference frequency, an equivalent time
can be computed.
Because the data and reference clock samplers are simultaneously strobed, a
precise clock phase is associated with each data sample. In a free-running mode,
the data samples are displayed relative to one complete 0-to-2π reference cycle.
The eye diagram produced differs from that which the scope ordinarily would
display without the phase reference module because each data point has been
corrected according to the phase measurement.
A triggered mode of operation also is supported by the 82A04. In this case, the
oscilloscope performs its usual sequential sampling action synchronized to a
repetitive trigger. The data and reference clock are simultaneously sampled at
the time the oscilloscope samples.
If the reference clock sampler is ignored for the moment, this mode mimics that
of a traditional sampling scope except that the actual data sampling is
performed by the 82A04. Samples are taken relative to the trigger just as they
would be under normal sampling-scope operation. The difference is that the
reference clock sampler is providing a separate, very accurate timing of the
scope time base.
In effect, each sample is tagged with the integer number of reference clock
cycles and the phase of the cycle in which the data was sampled. The reference
clock oscillator is being used to provide precise sample timing with greater
accuracy than the scope time base. This additional information allows the data
to be displayed avoiding scope time-base nonlinearities or jitter. The advantage
provided depends on the degree to which the reference clock is more stable and
accurate than the scope time base.
Another issue related to clock recovery is jitter bandwidth. Because of a
sampling scope’s several nanosecond delay from trigger to sample point, the
effect of high-frequency jitter on the sampled data may be very different from
its effect on the trigger some number of nanoseconds earlier. Worst case, if the
jitter has a component with a period similar to the trigger-to-sample delay, the
data can be acquired with twice the actual jitter.
The reference clock measurement approach used in the Tektronix 82A04 Module
avoids most of this problem by simultaneously strobing the data and reference
sampling gates. Nevertheless, even in this technique, the small mismatch between
the sampling gates themselves contributes similar types of errors albeit at much
higher frequency.
Another technique that is used in some clock recovery modules is to provide an
approximately equal delay in the data path. For example, in the Agilent 83496A
Optical Clock Recovery Option, a length of optical fiber is built in to align
the trigger event and sampling instant. Even though the compensation will not be
exact, it considerably minimizes the problem of constructive interference that
otherwise could exist.
Conclusion
As the speed of serial data communications links continues to increase,
distinctions between analog and digital technologies that are obvious at low
frequencies become blurred. Digital modulation schemes no longer produce only a
fixed set of allowed phase/magnitude constellation points, for example. Rather,
each ideal position has grown in width and height because of noise, crosstalk,
phase offset, gain errors, and various varieties of jitter. What kinds of test
tools best address these analog factors?
DSOs, the largest segment of the oscilloscope market by sales, continue to
advance the art of digitization while also achieving ever-greater bandwidths.
Nevertheless, performance is only just adequate for today’s serial data speeds.
To make the best measurements possible on high-speed signals, would you
voluntarily choose to include only the fifth harmonic and limit yourself to 8-b
resolution?
Sampling scopes have long addressed the acquisition of high-speed signals but
can be hard to use and slow to display the desired waveform. New digital
techniques teamed with the sampling scope’s high bandwidth and resolution have
resulted in instruments with very attractive features. And, because of automated
functions such as Agilent’s PatternLock and LeCroy’s CIS, complex acquisition
modes are greatly simplified.
In addition, sampling scopes recently have enjoyed increased interest as an
alternative to RF network analyzers. Time-domain reflectometry, making full use
of the sampling scope’s high resolution in both the amplitude and time axes, has
become a growth market.
S-parameters determined via scope-based TDR typically correlate to within a few
percent of network analyzer results. For many non-RF engineers, the network
analyzer’s slightly better performance is easily offset by its complex
calibration, high cost, and unfamiliar operation.
Today’s sampling scopes are not the same instruments that your grandfather used,
as the saying goes. They continue to exceed the bandwidth and sampling rate of
the best DSOs by a large margin but have gained many new capabilities as well,
such as jitter measurement and analysis. Why not evaluate a sampling scope
before starting your next serial data project? You may be pleasantly surprised.
Reference
1. “Ultra-Low Jitter Performance With Phase-Reference Module 82A04 and